| Coating/substrate combinations |
- more than 70 coating/substrate combinations (standard version)
- coatings on flat and curved surfaces
- coatings on small components and wires
- 10 preset types of metal: Cr, Ni, Cu, brass, Zn, Ag, Sn, Pb, Cd, Au
- 8 further metal coatings for special applications
- measuring range: 0.05…75 μm
|
|
| Measuring cell |
|
|
| Measuring surface |
- gasket 4 mm2/6.2 x 10-3 inch2
- mask 1 mm2/1.55 x 10-3 inch2
- mask 0.25 mm2/0.388 x 10-3 inch2 (deplated area nearly invisible)
- electrolyte cup 0.25 mm2 to approx. 16 mm2/(0.388–24.8) x 10-3 inch2 (optional)
|
|
| Adjustable settings for optimum results |
- maximum of 8 deplating speeds in the range of approx. 0.3–40 μm (0.012–1.57) mils/min
- directly adjustable calibration factors; individual setting according tometal type and measuring surface
- calibration setting with the aid of thickness standards
- variable shut-down speed to suppress interference or for measuring alloy zonesbetween coating and substrate
|
|
| Data storage with GalvanoTest |
- number of memories for various metals: 18
- number of readings which can be stored and evaluated: 2000
- non-volatile memory retains all calibration settings, readings and statistical valuesafter the gauge has been switched off
|
|
| Statistical evaluation |
- display of 6 statistical values – mean value, standard deviation, coefficient of variation,number of readings, highest and lowest single reading
- immediate or delayed display of statistical values
- immediate or delayed print-out of readings and statistics
- display and print-out of date and time – year, month, day, hour and minute
|
|
| GalvanoTest interfaces for peripheral devices |
- interface for MiniPrint data printer
- interface for an RS 232 C for connection to a PC
- analog output for connection to an x-t recorder to chart voltage curve
|
|
|
- Electrolyte alarm/saturation indicator
|
|
| Measuring uncertainty |
| 5 % of reading on a measuring area of 8 mm2 after calibration |
|
|
| Power supply |
| 110/220 V 50…60 Hz/10 Watt |
|
|
| Dimensions and weight |
| 260 mm x 250 mm x 100 mm/10.24 inch, 9.85 inch, 3.94 inch; approx. 2.5 kg |
| support with integrated pump approx. 3.0 kg |
|
|
| Features |
- print-out diagram showing measuring cell voltage levels over the coating; especially useful for detecting alloy zones between coating and substrate; with MiniPrint
- operator guide to the alphanumeric displays with language option:German, French or English
- unit of measurement in either metric (μm) or British system (mils)
- when switched on, the gauge displays details of the last readings
- illuminated voltmeter for monitoring the deplating process
- easy to use – clearly-designed controls and easy-to-follow operating instructionswith numerous examples
- wide range of extras for measurements on small components,wire and multi-layer coatings
|